AUTOCLAVE MICROJET

Product description

The Microjet autoclave is characterised by: - short medium preparation times thanks to the use of Direct Energy Transfer technology (12 minutes), - the ability to be installed anywhere due to its compact size and low weight, - ideal for laboratories where small quantities of media with varying compositions are required. Process parameters: Sterilisation temperature: 135°C Exposure time at sterilisation temperature: 80 s Heating time: 1.5–4.5 min Cooling time: approx. 4–8 minutes Total process time (from placing the vessel inside to removal) 7–12.5 minutes Temperature at the end of the process: 95 °C Technical specifications: Power supply 230 V/50 Hz Output power 0.95 kW Maximum current draw 8 A Operating pressure 3.6 bar Maximum pressure 5 bar Weight: 15 kg Process chamber volume: 2 l Electromagnetic wave frequency: 2450 MHz IP rating: IP 20 Data storage: MicroSD card